berbrazil.blogg.se

Digora optime dxr-60
Digora optime dxr-60










digora optime dxr-60
  1. #Digora optime dxr 60 serial
  2. #Digora optime dxr 60 manual
  3. #Digora optime dxr 60 software

Check that it goes to connector and that the slot on the pcb goes to the holder marked A. Replaced K board, calibrated and endurance test performed.Īfter testing, repair or adjustment, check that the unit is restored to the conditions necessary for normal use. Service terminal commands and functions are not explained in this manual.

#Digora optime dxr 60 software

Open Dental Software - Digora Soredex Optime Bridgeĭigora for Windows 2. Blue led at K board is used as the light source.

#Digora optime dxr 60 serial

Corrected from serial SL Service: The power adapter has accessible conductive part DC plug but this is newer opened or serviced and it is always measured when delivered from factory. Laser is faulty or laser connector is not connected. Attach the back cover If the cable reconnecting does not help, then ooptime user interface panel. Attach laser cable as in the image below. The laser can be activated if the light beam in the optic safety switch is blocked and the laser is switched on from the service terminal. Insert the main cover Open the covers and check that the cable is properly connected to K board and user interface board. Detach the connectors and remove the covers. Unit does not react to any other inputs than those given by the user through service terminal. Turn the unit on and start S2terminal The opime should now start without errors. Instead of the tool you can also use torque wrench. High voltage can be set on when the light beam of opto safety switch is blocked. It's a community-based project which helps to repair anything. Checking linear movement errors It is recommended first to check that the motor cable is connected to the motor and K board and that it is not broken motor is not rotating. However it is not always digoar display the error code. The -a flag tells Digora to change focus to itself.

#Digora optime dxr 60 manual

Double-click on Digora from DIGORA Optime DXR-60 Field Service Manual Uploader:Īfter calibration quit from service mode and turn the unit off and again on. Turn the unit on and start S2terminal This has been found as a root cause to many image quality related complaints with the new Optime. Check that there is no mechanical reason which prevents linear movement for example image plate or protective covers. They are instructed in the scanner firmware help which is accessible through service terminal.












Digora optime dxr-60